Manage data your way

Whether you’re connecting the tCOM+ or the Sentec Digital Monitor to other devices in your existing infrastructure, generating customized patient reports, or initiating remote monitoring, find out how Sentec enables safe and reliable data and connectivity solutions.

*Note: This is a global overview and may contain information on products not released in all markets/countries. Information is subject to change without notice.

tCOM-device-angled

Connectivity for the tCOM+

tCOM-device-angled
Patient Monitoring Systems (PMS)
Digistat® Smart Central

Required software versions

  • Digistat Suite 8.0 or higher
  • Sentec SMI Driver Version 1.1.10 or higher

Supported Sentec software versions

  • V1.03.00 or higher

Available parameters and data display

  • tcPCO₂
  • tcPO₂*
  • SpO₂
  • PR
  • Trend data display

*PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected

Sentec Interface Settings

Interface Options
Enable SDM Compatibility Mode

Encryption
Disable UDP Encryption

Firewall
Open SMI and SDP Ports in Firewall Settings

Additional accessories required

LAN cable

Carescape B450, Carescape B650, Carescape B850, Carescape Canvas

Required software versions

  • Carescape B450: V1 or higher
  • Carescape B650: V1 or higher
  • Carescape B850: V1 or higher
  • Carescape Canvas: V3.3 or higher

Supported Sentec software versions

  • V1.03.00 or higher

Available parameters and data display

  • tcPCO₂
  • tcPO₂*
  • SpO₂
  • PR
  • Trend data display

*PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected

Sentec interface settings

Serial Interface Protocol:
S
entecLink

Baud rate
:
19200

Interface Options:
Enable SDM Compatibility Mode

Encryption:
Disable Serial Encryption

Additional accessories required

Unity ID, DIDCA SENTEC, standard LAN cables (GE)

BeneView T5/T6/T8/T9
BeneVision N12/N15/N17/N19/N22

Required software versions

BeneView T5/T6/T8/T9:
5.25 or higher

BeneVision N12, N15, N17, N19, N22:
V2.0 or higher

Supported Sentec software versions

  • V1.03.00 or higher

Available parameters and data display

    • tcPCO
    • tcPO*
    • SpO
    • PR
    • Trend data display

*PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected

Sentec interface settings

Serial interface protocol:
SentecLink

Baud rate:
115200

Interface Options:
Enable SDM Compatibility Mode

Encryption:
Disable Serial Encryption

Additional accessories required

BeneLink ID Adapter and Cable C (REF PN-115-011468-00 Mindray)

Mindray installation manual

BeneVision Installation manual

Intellivue

Required software versions

IntelliVue MP 40/50/60/70/80/90 or MX600/700/800 via VueLink Interface:
All SW Versions

IntelliVue MP40/50/60/70/80/90 or MX600/700/750/800/850 with IntelliBridge EC10:
H.15 or higher

IntelliVue MX400/450/500/550 with IntelliBridge EC10:
All SW Versions

Supported Sentec software versions

IntelliVue MP 40/50/60/70/80/90 or MX600/700/800/850 via VueLink Interface:

  • V 1.03.00 or higher

IntelliVue MP40/50/60/70/80/90 or MX600/700/750/800 with IntelliBridge EC10:

  • V 1.03.00 or higher

IntelliVue MX400/450/500/550 with IntelliBridge EC10:

  • V 1.03.00 or higher

Available parameters and data display

  • tcPCO₂
  • tcPO₂*
  • SpO₂
  • PR

* PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected.

Sentec interface settings

Serial Interface Protocol:
Philips Vuelink/Intellibridge 1/2

Additional accessories required

IntelliBridge EC5 and EC10, LAN cable (Philips)

Philips guides and manuals

Patient Data Management Systems (PDMS)
Medical Device information platform (MDIP)

Required software versions

  • Sentec A5.x and higher

Supported Sentec software versions

  • V1.03.00 or higher

Available parameters and data display

  • tcPCO₂
  • tcPO₂*
  • SpO₂
  • PR

* PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected.

Sentec interface settings

Serial Interface Protocol:
SentecLink

Baud rate
115200 or lower

Interface Options:
Enable SDM Compatibility Mode

Encryption:
Disable Serial Encryption

Additional accessories required

Device ID Module (DIM )(Capsule), Standard RS232 and LAN cable

INTEGRATED CARE MANAGER (ICM)

Required software versions

  • Version 9.2

Supported Sentec software versions

  • V1.03.00 or higher

Available parameters and data display

  • tcPCO₂
  • tcPO*
  • SpO₂
  • PR

* PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected.

Sentec interface settings

Serial Interface Protocol:
SentecLink

Baud rate
115200 or lower

Interface Options:
Enable SDM Compatibility Mode

Encryption:
Disable Serial Encryption

Additional accessories required

Standard RS232 cable or USB2Serial converter cable

VISI-DOWNLOAD

Required software versions

  • 140715 or higher

Supported Sentec software versions

  • V01.03.00

Available parameters and data display

  • tcPCO₂
  • tcPCO₂ (drift corrected)
  • tcPO₂*
  • SpO₂
  • PR
  • Trend data display

*PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected.

Sentec interface settings

Serial Interface Protocol:
SentecLink

Baud rate:
115200 or lower

Interface Options:
Enable SDM Compatibility Mode

Encryption:
Disable Serial Encryption

Additional accessories required

Standard RS232 cable or USB2 Serial converter cable

Ventilators
VENDOM 30/40

Required software versions

  • V 2.05 or higher

Supported Sentec software versions

  • V1.03.00 or higher

Available parameters and data display

  • tcPCO₂
  • SpO₂
  • PR
  • Trend data display

Sentec interface settings

Serial Interface Protocol:
SentecLink

Baud rate:
115200

Interface Options:
Enable SDM Compatibility Mode

Encryption:
Disable Serial Encryption

Additional accessories required

USB2 Serial converter cable

Vivo 45, Nippy 4+

Required software versions

Vivo 45:
1.0.33 or higher

Vivo 45 LS:
3.0.44 or higher

Vivo 45 L (US Only):
5.0.01 or higher

Nippy 4+ (UK Only):
4.0.51 or higher

Supported Sentec software versions

  • V1.03.00 or higher

Available parameters and data display

  • tcPCO
  • PR
  • Trend data display

Sentec interface settings

Serial Interface Protocol:
SentecLink

Baud rate:
57600 or lower

Interface Options:
Enable SDM Compatibility Mode

Encryption:
Disable Serial Encryption

Additional accessories required

Serial adapter cable (REF 006179) (Breas)

EO-150

Required software versions

  • V 5.0.0 or higher

Supported Sentec software versions

  • V1.03.00 or higher

Available parameters and data display

  • tcPCO
  • tcPCO (drift corrected)
  • SpO
  • PR
  • Trend data display

Sentec Interface settings

Serial Interface Protocol:
SentecLink

Baud rate:
115200

Interface Options:
Enable SDM Compatibility Mode

Encryption:
Disable Serial Encryption

Additional accessories required

Standard serial RS232 to USB converter cable, optional: EOVE Clinical Software

Triology, A30, A40

Required software versions

Trilogy Evo, Trilogy Evo O2, Trilogy EV300:

  • 1.05.06.00 or higher (US)
  • 1.06.06.00 or higher (outside US)

A30 EFL, A40 Pro, A40 EFL:

  • 1.1.0.0 and higher

Supported Sentec software versions

V.01.03.00 or higher

Available parameters and data display

Trilogy Evo, Trilogy Evo O2, Trilogy EV300:

  • tcPCO
  • Trend data display

A30 EFL, A40 Pro, A40 EFL:

  • tcPCO
  • SpO
  • PR
  • Trend data display

Sentec Interface Settings

Serial Interface Protocol:

Trilogy Evo, Trilogy Evo O2, Trilogy EV300:

  • Serial Interface Protocol: SentecLink
  • Baud rate: 19200

A30 EFL, A40 Pro, A40 EFL:

  • Serial Interface Protocol: SentecLink
  • Baud rate: 115200

Interface Options:
Enable SDM Compatibility Mode

Encryption:
Disable Serial Encryption

Additional accessories required

Trilogy Evo, Trilogy Evo O2, Trilogy EV300:
Isolated USB-A to DB9 cable (REF 1127405), serial null modem cable (Philips Respironics)

A30 EFL, A40 Pro, A40 EFL:

Transcutaneous CO₂ Connection Cable (REF 1146885), Link Module (REF 1061644A), Philips Null Modem, SD Card (Philips Respironics)

Additional information for operation:

Trilogy Evo, Trilogy Evo O2, Trilogy EV300:
Please ensure that the tCOM+ monitor is started and ‘Ready for use’ before connecting with the Trilogy ventilator. The required tCOM+ profile must be selected on tCOM+ before connecting with Trilogy ventilators. 

 

Monsson 4 via remote display:

Required software versions

  • Monsoon 4: 1.0 or higher
  • Remote Display: 1.0 or higher

Supported Sentec software versions

  • V01.03.00

Available parameters and data display

  • tcPCO₂
  • SpO₂
  • PR
  • Trend data display

Sentec interface settings

Serial Interface Protocol:
SentecLink

Baud rate
115200 or lower

Interface Options:
Enable SDM Compatibility Mode

Encryption:
Disable Serial Encryption

Additional accessories required

ACUDis remote display, RJ45 to RS232 cable (Thora Tech)

Digital Connectivity to PG/PSG
NOX A1, NOX A1s, NOX C1

Required software versions

NOX A1 (via W7Link):
FW 1.1.6 and Noxturnal 5.1.3

NOX A1s (via C1):
FW 3.0.9 and Noxturnal 6.3.1 or higher

NOX C1:
FW 1.1.0 and Noxturnal 6.0.0 or higher

Supported Sentec software versions

  • V1.03.00 or higher

Available parameters and data display

  • tcPCO
  • tcPO*
  • SpO
  • PR
  • Trend data display

*PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected.

Sentec Interface Settings

Serial Integration (C1 and W7LInk)

Serial Interface Protocol
SentecLink

Baud rate
115200

Interface Options:
Enable SDM Compatibility Mode

Encryption:
Disable Serial Encryption

Lan Integration (C1)

Interface Options:
Enable SDM Compatibility Mode

Encryption:
Disable LAN Encryption

Firewall:
Open SMI and SDP Ports in Firewall Settings

Additional accessories required

NOX A1 (via W7Link)
W7Link, Mini USB Cable , Noxturnal (Nox)

NOX C1 (serial)
RS-232 to 3.5mm jack serial cable, Standard Ethernet cable, Noxturnal (Nox)

NOX C1 (LAN)
Switch, three standard Ethernet cables, Noxturnal (Nox)

Connectivity for the Sentec Digital Monitor

Patient Monitoring Systems (PMS)
Digistat® Smart Central

Required software versions

  • Digistat Suite 8.0 or higher
  • Sentec SMI Driver Version 1.1.10 or higher

Supported Sentec software versions

  • V08.04.x
  • V08.05.x

Available parameters and data display

  • tcPCO₂
  • tcPO₂*
  • SpO₂
  • PR
  • Trend data display

*PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected

Sentec Interface Settings

LAN

Additional accessories required

LAN cable

ASCOM guides and manuals

ASCOM connectivity guide

Infinity® Delta, Infinity® Delta XL, Infinity® Kappa, Infinity® Omega Series:

Supported Sentec software versions

  • V07.xx.x
  • V08.00.x
  • V08.01.x
  • V08.02.x
  • V08.03.x
  • V08.04.x
  • V08.05.x

Available parameters and data display

  • tcPCO₂
  • tcPO₂

*PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected

Sentec Interface Settings

Serial Interface Protocol:
TCB-Protocol

Additional accessories required

MIB Connector, MIB II Converter, Infinity Adapter, Patch cable, Standard RS232 cable (Dräger)

Dräger guides and manuals

Carescape B450, Carescape B650, Carescape B850, Carescape Canvas, Solar 8000 M, Solar 8000 i, Dash 5000:

Required software versions

  • Carescape B450: V1 or higher
  • Carescape B650: V1 or higher
  • Carescape B850: V1 or higher
  • Carescape Canvas: V3.3 or higher
  • Solar 8000 M: V4A
  • Solar 8000 i: V4E
  • Dash 5000: V3A

Supported Sentec software versions

  • V08.00.x
  • V08.01.x
  • V08.02.x
  • V08.03.x
  • V08.04.x
  • V08.05.x

Available parameters and data display

  • tcPCO₂
  • tcPO₂*
  • SpO₂
  • PR
  • Trend data display

*PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected

Sentec Interface Settings

Serial Interface Protocol:
S
entecLink
Baud rate:
19200

Additional accessories required

Unity ID, DIDCA SENTEC, standard LAN cables (GE)

GE Healthcare guides and manuals

GE Connectivity Guide

BeneView T5/T6/T8/T9
BeneVision N12/N15/N17/N19/N22

Required software versions

BeneView T5/T6/T8/T9:
5.25 or higher

BeneVision N12, N15, N17, N19, N22:
V2.0 or higher

Supported Sentec software versions

  • V08.00.x
  • V08.01.x
  • V08.02.x
  • V08.03.x
  • V08.04.x
  • V08.05.x

Available parameters and data display

  • tcPCO₂
  • tcPO₂*
  • SpO₂
  • PR
  • Trend data display

*PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected

Sentec Interface Settings

Serial Interface Protocol:
SentecLink
Baud rate:
115200

Additional accessories required

BeneLink ID Adapter and Cable C (REF PN-115-011468-00 Mindray)

Mindray guides and manuals

Intellivue

Required software versions

IntelliVue MP 40/50/60/70/80/90 or MX600/700/800 via VueLink Interface:
All SW Versions

IntelliVue MP40/50/60/70/80/90 or MX600/700/750/800/850 with IntelliBridge EC10:
H.15 or higher

IntelliVue MX400/450/500/550 with IntelliBridge EC10:
All SW Versions

Supported Sentec software versions

IntelliVue MP 40/50/60/70/80/90 or MX600/700/800/850 via VueLink Interface:

  • V07.xx.x
  • V08.00.x
  • V08.01.x
  • V08.02.x
  • V08.03.x
  • V08.04.x
  • V08.05.x

IntelliVue MP40/50/60/70/80/90 or MX600/700/750/800 with IntelliBridge EC10:

  • V07.xx.x
  • V08.00.x
  • V08.01.x
  • V08.02.x
  • V08.03.x
  • V08.04.x
  • V08.05.x

IntelliVue MX400/450/500/550 with IntelliBridge EC10:

  • V07.xx.x
  • V08.00.x
  • V08.01.x
  • V08.02.x
  • V08.03.x
  • V08.04.x
  • V08.05.x

Available parameters and data display

  • tcPCO₂
  • tcPO₂*
  • SpO₂
  • PR

* PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected.

Sentec Interface Settings

Serial Interface Protocol:
Philips Vuelink/Intellibridge 1/2

Additional accessories required

IntelliBridge EC5 and EC10, LAN cable (Philips)

Philips guides and manuals

Flexport System UltraView SL (90442A-03)

Required software version

V2.00.02EN

Supported Sentec software version

  • V07.xx.x
  • V08.00.x
  • V08.01.x
  • V08.02.x
  • V08.03.x
  • V08.04.x
  • V08.05.x

Available parameters and data display

  • tcPCO
  • SpO
  • PR
  • Trend data display

Sentec Interface Settings

Serial Interface Protocol:
Spacelabs Flexport

Additional accessories required

Flexport interface, SDLC cable, SDLC terminator, Transition connector, Modular Cable (Spacelabs)

Spacelabs guides and manuals

Patient Data Management Systems (PDMS)
Medical Device information platform (MDIP)

Required software versions

  • Sentec A5.x and higher

Supported Sentec software versions

  • V07.xx.x
  • V08.00.x
  • V08.01.x
  • V08.02.x
  • V08.03.x
  • V08.04.x
  • V08.05.x

Available parameters and data display

  • tcPCO₂
  • tcPO₂*
  • SpO₂
  • PR

* PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected.

Sentec Interface Settings

Serial Interface Protocol:
SentecLink

Baud rate
115200 or lower

Additional accessories required

Device ID Module (DIM )(Capsule), Standard RS232 and LAN cable

Capsule guides and manuals

Capsule Connectivity Guide

INTEGRATED CARE MANAGER (ICM)

Required software versions

  • Version 9.2

Supported Sentec software versions

  • V08.00.x
  • V08.01.x
  • V08.02.x
  • V08.03.x
  • V08.04.x
  • V08.05.x

Available parameters and data display

  • tcPCO₂
  • TcPO₂*
  • SpO₂
  • PR

* PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected.

Sentec Interface Settings

Serial Interface Protocol:
SentecLink

Baud rate
115200 or lower

Additional accessories required

Standard RS232 cable or USB2Serial converter cable

Dräger guides and manuals

Dräger Integrated Care Manager Connectivity Guide

Mirrel

Required software versions

  • 1.0.2.0 or higher (driver)

Supported Sentec software versions

  • V08.05.x

Available parameters and data display

  • tcPCO₂
  • TcPO₂*
  • SPO₂
  • PR
  • HP
  • PI
  • Sensor temperature

*PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected.

Sentec Interface Settings

Serial integration
SentecLink

Baud rate
115200

Regional Availability

Japan Only

Metavision

Required software versions

  • V5 or higher

Supported Sentec software versions

  • V08.00.x
  • V08.01.x
  • V08.02.x
  • V08.03.x
  • V08.04.x
  • V08.05.x

Available parameters and data display

  • tcPCO₂
  • TcPO₂*
  • SPO₂
  • PR

*PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected.

Sentec Interface Settings

Serial Interface Protocol
SentecLink

Baud rate
115200

Additional accessories required

Digit Connect (iMDSoft), Serial to LAN converter cable 250372-15D, Standard LAN cable

iMDsoft guides and manuals

iMDsoft Connectivity Guide

Prime Gaia

Required software versions

  • 03-01 or higher

Supported Sentec software versions

  • V08.02.x
  • V08.03.x
  • V08.04.x
  • V08.05.x

Available parameters and data display

  • tcPCO₂
  • TcPO₂*
  • SpO₂
  • PR
  • HP
  • PI
  • Sensor temperature
  • Alarms and status codes

*PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected.

Sentec Interface Settings

  • Serial integration: SentecLink, baud rate 115200 or lower or alternative
  • Lan integration

Regional Availability

Japan Only

VISI-DOWNLOAD

Required software versions

  • 140715 or higher

Supported Sentec software versions

  • V08.00.x
  • V08.01.x
  • V08.02.x
  • V08.03.x
  • V08.04.x
  • V08.05.x

Available parameters and data display

  • tcPCO₂
  • tcPCO₂ (drift corrected)
  • TcPO₂*
  • SpO₂
  • PR
  • Trend data display

*PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected.

Sentec Interface Settings

Serial Interface Protocol:
SentecLink

Baud rate:
115200 or lower

Additional accessories required

Standard RS232 cable or USB2 Serial converter cable

Stowood guides and manuals

Stowood Connectivity Guide

TZM (via) UMG to GE Healthcare’s patient monitors

Required software versions

  • V1.0

Supported Sentec software versions

  • V08.00.x
  • V08.01.x
  • V08.02.x

Available parameters and data display

  • tcPCO
  • TcPO₂*

*PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected.

Sentec Interface Settings

Serial Interface Protocol:
SentecLink

Baud rate:
115200

Additional accessories required

UMG

NivolisAnton

Required software versions

  • SW-Version A009-074B0-4.4.1 or higher

Supported Sentec software versions

  • V08.00.x
  • V08.01.x
  • V08.02.x
  • V08.03.x
  • V08.04.x
  • V08.05.x

Available parameters and data display

  • tcPCO
  • tcPCO₂ Drift Corrected
  • SpO
  • PR
  • PI
  • Status codes
  • Signal quality
  • Download of SDL files for import into V-STATS (trend data display, analysis and reporting)

Sentec Interface Settings

Serial Interface Protocol:
SentecLink

Baud rate:
115200

Additional accessories required

NivolisAnton, screwdriver

Ventilators
VENDOM 30/40

Required software versions

  • V 2.05 or higher

Supported Sentec software versions

  • V07.xx.x
  • V08.00.x
  • V08.01.x
  • V08.02.x
  • V08.03.x
  • V08.04.x
  • V08.05.x

Available parameters and data display

  • tcPCO₂
  • SpO₂
  • PR
  • Trend data display

Sentec Interface Settings

Serial Interface Protocol:
SentecLink

Baud rate:
115200

Additional accessories required

USB2 Serial converter cable

Air Liquide guides and manuals

Air Liquide Connectivity Guide

Vivo 45, Nippy

Required software versions

Vivo 45:
1.0.33 or higher

Vivo 45 LS:
3.0.44 or higher

Vivo 45 L (US):
5.0.01 or higher

Nippy 4+ (UK only):
4.0.51 or higher

Supported Sentec software versions

  • V07.xx.x
  • V08.00.x
  • V08.01.x
  • V08.02.x
  • V08.03.x
  • V08.04.x

Available parameters and data display

  • tcPCO
  • PR
  • Trend data display

Sentec Interface Settings

Serial Interface Protocol:
SentecLink

Baud rate:
57600 or lower

Additional accessories required

Serial adapter cable (REF 006179) (Breas)

Breas guides and manuals

Breas Connectivity Guide

EO-150

Required software versions

  • V 5.0.0 or higher

Supported Sentec software versions

  • V08.00.x
  • V08.01.x
  • V08.02.x
  • V08.03.x
  • V08.04.x
  • V08.05.x

Available parameters and data display

  • tcPCO
  • tcPCO (drift corrected)
  • SpO
  • PR
  • Trend data display

Sentec Interface Settings

Serial Interface Protocol:
SentecLink

Baud rate:
115200

Additional accessories required

Standard serial RS232 to USB converter cable, optional: EOVE Clinical Software

EOVE guides and manuals Download the EOVE guide:

Eove Connectivity Guide

Trilogy, A30, A40

Required software versions

Trilogy Evo, Trilogy Evo O2, Trilogy EV300:

  • 1.05.06.00 or higher (US)
  • 1.06.06.00 or higher (outside US)

A30 EFL, A40 Pro, A40 EFL:

  • 1.1.0.0 and higher:

Supported Sentec software versions

  • V08.00.x
  • V08.01.x
  • V08.02.x
  • V08.03.x
  • V08.04.x
  • V08.05.x

Available parameters and data display

Trilogy Evo, Trilogy Evo O2, Trilogy EV300:

  • tcPCO
  • Trend data display

A30 EFL, A40 Pro, A40 EFL:

  • tcPCO
  • SpO
  • PR
  • Trend data display

Sentec Interface Settings

Trilogy Evo, Trilogy Evo O2, Trilogy EV300:

  • Serial Interface Protocol: SentecLink
  • Baud rate: 19200

A30 EFL, A40 Pro, A40 EFL:

  • Serial Interface Protocol: SentecLink
  • Baud rate: 115200

Additional accessories required

Trilogy Evo, Trilogy Evo O2, Trilogy EV300:
Isolated USB-A to DB9 cable (REF 1127405), serial null modem cable (Philips Respironics)

A30 EFL, A40 Pro and A40 EFL:
Transcutaneous CO Connection Cable (REF 1146885), Link Module (REF 1061644A), Philips Null Modem, SD Card (Philips Respironics)

Philips guides and manuals

Philips Trilogy Evo Connectivity Guide

Monsson 4 via remote display:

Required software versions

  • Monsoon 4: 1.0 or higher
  • Remote Display: 1.0 or higher

Supported Sentec software versions

  • V08.00.x
  • V08.01.x
  • V08.02.x
  • V08.03.x
  • V08.04.x
  • V08.05.x

Available parameters and data display

  • tcPCO₂
  • SpO₂
  • PR
  • Trend data display

Sentec Interface Settings

Serial Interface Protocol:
SentecLink

Baud rate
115200 or lower

Additional accessories required

ACUDis remote display, RJ45 to RS232 cable (Thora Tech)

Thora Tech guides and manuals

Thora Tech Connectivity Guide

Digital Connectivity to PG/PSG
CID102L, CID102LB, CID-LXe, CID-LXe-206d, CID-LX, CID-LX-206d

Required software versions

2.2.4 or higher

Supported software versions

  • V07.xx.x
  • V08.00.x
  • V08.01.x
  • V08.02.x
  • V08.03.x
  • V08.04.x
  • V08.05.x

Available parameters and data display

  • tcPCO
  • tcPCO (drift corrected)
  • SpO
  • PR
  • Trend data display

Sentec Interface Settings

Serial Interface Protocol
SentecLink

Baud rate
115200

Additional accessories required?

Cidelec Analysis Software, Standard RS232 cable or USB2Serial converter cable

Cidelec guides and manuals

Cidelec Connectivity Guide

NOX A1, NOX A1s, NOX C1, NOX T3

Required software versions

NOX A1 (via W7Link):
FW 1.1.6 and Noxturnal 5.1.3

NOX A1s (via C1):
FW 3.0.9 and Noxturnal 6.3.1 or higher

NOX C1:
FW 1.1.0 and Noxturnal 6.0.0 or higher

NOX T3 (via W7Link):
FW 1.6.1 and Noxturnal 5.1.3

Supported Sentec software versions

  • V08.00.x
  • V08.01.x
  • V08.02.x
  • V08.03.x
  • V08.04.x
  • V08.05.x

Available parameters and data display

  • tcPCO
  • tcPO*
  • SpO
  • PR
  • Trend data display

*PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected.

Sentec Interface Settings

Serial Interface Protocol
SentecLink

Baud rate
115200

NOX C1:
LAN integration possible (alternative)

Additional accessories required

NOX A1 or T3 (via W7Link)
W7Link, Mini USB Cable , Noxturnal (Nox)

NOX C1 (serial)
RS-232 to 3.5mm jack serial cable, Standard Ethernet cable, Noxturnal (Nox)

NOX C1 (LAN)
Switch, three standard Ethernet cables, Noxturnal (Nox)

Nox guides and manuals

Nox Medical Connectivity Guide

Analog Connections for the tCOM+ and the Sentec Digital Monitor

SomnoStar Pro: PSG Cable E
CID102L8-D: PSG Cable P
  • E-Series: PSG Cable E
  • Grael: PSG Cable E
  • Grael 4k: PSG Cable E
  • Siesta: PSG Cable G
  • Somte: PSG Cable E
  • SleepDoc Porti 7: PSG Cable D
  • SleepDoc Porti 8: PSG Cable D
  • MiniScreen PRO: PSG Cable D
  • MiniScreen plus: PSG Cable D
  • Sonata: PSG Cable B
  • Comet / Aurora: PSG Cable E
  • Comlab: PSG Cable H
  • Deltamed Brain Box: PSG Cable J
  • Deltamed Duo 44: PSG Cable H
  • Embletta Gold: PSG Cable L
  • Embletta MPR PG: PSG Cable L
  • Embla NDx: PSG Cable F
  • Embla SDx: PSG Cable E
  • Embla S7000 / N7000: PSG Cable A
  • Embla Titanium (Isolation Box): PSG Cable E
  • Embla Titanium: PSG Cable K
  • Embla Installation Manual
  • Mdrive: PSG Cable N
  • TX Proxy: PSG Cable N
  • PAIB2 interface box: PSG Cable F
  • Sleepvirtual: PSG Cable E
  • BWII (SI1089)**: PSG Cable E
  • BWIII PSG: Cable E

** Product number of accessory cables or optocoupler required for installation in brackets behind the product name

  • SOMNOscreen (SSC082)**: PSG Cable Q
  • SOMNOscreen plus (SSC082)**: PSG Cable Q

** Product number of accessory cables or optocoupler required for installation in brackets behind the product name

  • Visi-3: PSG Cable E
  • Visi-System: PSG Cable E
  • Alice 3 / 4 / 5 / 6 (square plug): PSG Cable B
  • Alice 3 / 4 / 5 / 6 (round plug): PSG Cable F
  • SOMNOlab: PSG Cable C
  • SOMNOlab2: PSG Cable C

Please refer to the tCOM+ menu ‘Advanced Settings/Interfaces/Analog/PSG/Channel Assignment’ respectively the Sentec Digital Monitor (SDM) menu ‘Interfaces/Analog Outputs’ for ‘Channel-to-Parameter’ assignment.

For instructions to interface the tCOM+ or SDM with polygraphic or polysomnographic systems please refer to the respective manufacturer’s instructions for use, to the tCOM+ IFU (HBQ-179), to the chapter ‘Analog Outputs’ in the Technical Manual for the SDM (HB-005752 – HBQ-11) or the Analog Interface Calibration Guide for the SDM (RF-013071- MMQ-96).

The IFUs and Technical Manual are available as a PDF on the Sentec website at https://www.sentec.com/ifu