Manage data your way
Whether you’re connecting the tCOM+ or the Sentec Digital Monitor to other devices in your existing infrastructure, generating customized patient reports, or initiating remote monitoring, find out how Sentec enables safe and reliable data and connectivity solutions.
*Note: This is a global overview and may contain information on products not released in all markets/countries. Information is subject to change without notice.
Connectivity for the tCOM+
Patient Monitoring Systems (PMS)
Digistat® Smart Central
Required software versions
- Digistat Suite 8.0 or higher
- Sentec SMI Driver Version 1.1.10 or higher
Supported Sentec software versions
- V1.03.00 or higher
Available parameters and data display
- tcPCO₂
- tcPO₂*
- SpO₂
- PR
- Trend data display
*PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected
Sentec Interface Settings
Interface Options
Enable SDM Compatibility Mode
Encryption
Disable UDP Encryption
Firewall
Open SMI and SDP Ports in Firewall Settings
Additional accessories required
LAN cable
Carescape B450, Carescape B650, Carescape B850, Carescape Canvas
Required software versions
- Carescape B450: V1 or higher
- Carescape B650: V1 or higher
- Carescape B850: V1 or higher
- Carescape Canvas: V3.3 or higher
Supported Sentec software versions
- V1.03.00 or higher
Available parameters and data display
- tcPCO₂
- tcPO₂*
- SpO₂
- PR
- Trend data display
*PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected
Sentec interface settings
Serial Interface Protocol:
SentecLink
Baud rate:
19200
Interface Options:
Enable SDM Compatibility Mode
Encryption:
Disable Serial Encryption
Additional accessories required
Unity ID, DIDCA SENTEC, standard LAN cables (GE)
BeneView T5/T6/T8/T9
BeneVision N12/N15/N17/N19/N22
Required software versions
BeneView T5/T6/T8/T9:
5.25 or higher
BeneVision N12, N15, N17, N19, N22:
V2.0 or higher
Supported Sentec software versions
- V1.03.00 or higher
Available parameters and data display
-
- tcPCO₂
- tcPO₂*
- SpO₂
- PR
- Trend data display
*PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected
Sentec interface settings
Serial interface protocol:
SentecLink
Baud rate:
115200
Interface Options:
Enable SDM Compatibility Mode
Encryption:
Disable Serial Encryption
Additional accessories required
BeneLink ID Adapter and Cable C (REF PN-115-011468-00 Mindray)
Mindray installation manual
Intellivue
Required software versions
IntelliVue MP 40/50/60/70/80/90 or MX600/700/800 via VueLink Interface:
All SW Versions
IntelliVue MP40/50/60/70/80/90 or MX600/700/750/800/850 with IntelliBridge EC10:
H.15 or higher
IntelliVue MX400/450/500/550 with IntelliBridge EC10:
All SW Versions
Supported Sentec software versions
IntelliVue MP 40/50/60/70/80/90 or MX600/700/800/850 via VueLink Interface:
- V 1.03.00 or higher
IntelliVue MP40/50/60/70/80/90 or MX600/700/750/800 with IntelliBridge EC10:
- V 1.03.00 or higher
IntelliVue MX400/450/500/550 with IntelliBridge EC10:
- V 1.03.00 or higher
Available parameters and data display
- tcPCO₂
- tcPO₂*
- SpO₂
- PR
* PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected.
Sentec interface settings
Serial Interface Protocol:
Philips Vuelink/Intellibridge 1/2
Additional accessories required
IntelliBridge EC5 and EC10, LAN cable (Philips)
Philips guides and manuals
Patient Data Management Systems (PDMS)
Medical Device information platform (MDIP)
Required software versions
- Sentec A5.x and higher
Supported Sentec software versions
- V1.03.00 or higher
Available parameters and data display
- tcPCO₂
- tcPO₂*
- SpO₂
- PR
* PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected.
Sentec interface settings
Serial Interface Protocol:
SentecLink
Baud rate
115200 or lower
Interface Options:
Enable SDM Compatibility Mode
Encryption:
Disable Serial Encryption
Additional accessories required
Device ID Module (DIM )(Capsule), Standard RS232 and LAN cable
INTEGRATED CARE MANAGER (ICM)
Required software versions
- Version 9.2
Supported Sentec software versions
- V1.03.00 or higher
Available parameters and data display
- tcPCO₂
- tcPO₂*
- SpO₂
- PR
* PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected.
Sentec interface settings
Serial Interface Protocol:
SentecLink
Baud rate
115200 or lower
Interface Options:
Enable SDM Compatibility Mode
Encryption:
Disable Serial Encryption
Additional accessories required
Standard RS232 cable or USB2Serial converter cable
VISI-DOWNLOAD
Required software versions
- 140715 or higher
Supported Sentec software versions
- V01.03.00
Available parameters and data display
- tcPCO₂
- tcPCO₂ (drift corrected)
- tcPO₂*
- SpO₂
- PR
- Trend data display
*PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected.
Sentec interface settings
Serial Interface Protocol:
SentecLink
Baud rate:
115200 or lower
Interface Options:
Enable SDM Compatibility Mode
Encryption:
Disable Serial Encryption
Additional accessories required
Standard RS232 cable or USB2 Serial converter cable
Ventilators
VENDOM 30/40
Required software versions
- V 2.05 or higher
Supported Sentec software versions
- V1.03.00 or higher
Available parameters and data display
- tcPCO₂
- SpO₂
- PR
- Trend data display
Sentec interface settings
Serial Interface Protocol:
SentecLink
Baud rate:
115200
Interface Options:
Enable SDM Compatibility Mode
Encryption:
Disable Serial Encryption
Additional accessories required
USB2 Serial converter cable
Vivo 45, Nippy 4+
Required software versions
Vivo 45:
1.0.33 or higher
Vivo 45 LS:
3.0.44 or higher
Vivo 45 L (US Only):
5.0.01 or higher
Nippy 4+ (UK Only):
4.0.51 or higher
Supported Sentec software versions
- V1.03.00 or higher
Available parameters and data display
- tcPCO₂
- PR
- Trend data display
Sentec interface settings
Serial Interface Protocol:
SentecLink
Baud rate:
57600 or lower
Interface Options:
Enable SDM Compatibility Mode
Encryption:
Disable Serial Encryption
Additional accessories required
Serial adapter cable (REF 006179) (Breas)
EO-150
Required software versions
- V 5.0.0 or higher
Supported Sentec software versions
- V1.03.00 or higher
Available parameters and data display
- tcPCO₂
- tcPCO₂ (drift corrected)
- SpO₂
- PR
- Trend data display
Sentec Interface settings
Serial Interface Protocol:
SentecLink
Baud rate:
115200
Interface Options:
Enable SDM Compatibility Mode
Encryption:
Disable Serial Encryption
Additional accessories required
Standard serial RS232 to USB converter cable, optional: EOVE Clinical Software
Triology, A30, A40
Required software versions
Trilogy Evo, Trilogy Evo O2, Trilogy EV300:
- 1.05.06.00 or higher (US)
- 1.06.06.00 or higher (outside US)
A30 EFL, A40 Pro, A40 EFL:
- 1.1.0.0 and higher
Supported Sentec software versions
V.01.03.00 or higher
Available parameters and data display
Trilogy Evo, Trilogy Evo O2, Trilogy EV300:
- tcPCO₂
- Trend data display
A30 EFL, A40 Pro, A40 EFL:
- tcPCO₂
- SpO₂
- PR
- Trend data display
Sentec Interface Settings
Serial Interface Protocol:
Trilogy Evo, Trilogy Evo O2, Trilogy EV300:
- Serial Interface Protocol: SentecLink
- Baud rate: 19200
A30 EFL, A40 Pro, A40 EFL:
- Serial Interface Protocol: SentecLink
- Baud rate: 115200
Interface Options:
Enable SDM Compatibility Mode
Encryption:
Disable Serial Encryption
Additional accessories required
Trilogy Evo, Trilogy Evo O2, Trilogy EV300:
Isolated USB-A to DB9 cable (REF 1127405), serial null modem cable (Philips Respironics)
A30 EFL, A40 Pro, A40 EFL:
Transcutaneous CO₂ Connection Cable (REF 1146885), Link Module (REF 1061644A), Philips Null Modem, SD Card (Philips Respironics)
Additional information for operation:
Trilogy Evo, Trilogy Evo O2, Trilogy EV300:
Please ensure that the tCOM+ monitor is started and ‘Ready for use’ before connecting with the Trilogy ventilator. The required tCOM+ profile must be selected on tCOM+ before connecting with Trilogy ventilators.
Monsson 4 via remote display:
Required software versions
- Monsoon 4: 1.0 or higher
- Remote Display: 1.0 or higher
Supported Sentec software versions
- V01.03.00
Available parameters and data display
- tcPCO₂
- SpO₂
- PR
- Trend data display
Sentec interface settings
Serial Interface Protocol:
SentecLink
Baud rate
115200 or lower
Interface Options:
Enable SDM Compatibility Mode
Encryption:
Disable Serial Encryption
Additional accessories required
ACUDis remote display, RJ45 to RS232 cable (Thora Tech)
Digital Connectivity to PG/PSG
NOX A1, NOX A1s, NOX C1
Required software versions
NOX A1 (via W7Link):
FW 1.1.6 and Noxturnal 5.1.3
NOX A1s (via C1):
FW 3.0.9 and Noxturnal 6.3.1 or higher
NOX C1:
FW 1.1.0 and Noxturnal 6.0.0 or higher
Supported Sentec software versions
- V1.03.00 or higher
Available parameters and data display
- tcPCO₂
- tcPO₂*
- SpO₂
- PR
- Trend data display
*PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected.
Sentec Interface Settings
Serial Integration (C1 and W7LInk)
Serial Interface Protocol
SentecLink
Baud rate
115200
Interface Options:
Enable SDM Compatibility Mode
Encryption:
Disable Serial Encryption
Lan Integration (C1)
Interface Options:
Enable SDM Compatibility Mode
Encryption:
Disable LAN Encryption
Firewall:
Open SMI and SDP Ports in Firewall Settings
Additional accessories required
NOX A1 (via W7Link)
W7Link, Mini USB Cable , Noxturnal (Nox)
NOX C1 (serial)
RS-232 to 3.5mm jack serial cable, Standard Ethernet cable, Noxturnal (Nox)
NOX C1 (LAN)
Switch, three standard Ethernet cables, Noxturnal (Nox)
Connectivity for the Sentec Digital Monitor
Patient Monitoring Systems (PMS)
Digistat® Smart Central
Required software versions
- Digistat Suite 8.0 or higher
- Sentec SMI Driver Version 1.1.10 or higher
Supported Sentec software versions
- V08.04.x
- V08.05.x
Available parameters and data display
- tcPCO₂
- tcPO₂*
- SpO₂
- PR
- Trend data display
*PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected
Sentec Interface Settings
LAN
Additional accessories required
LAN cable
ASCOM guides and manuals
Infinity® Delta, Infinity® Delta XL, Infinity® Kappa, Infinity® Omega Series:
Supported Sentec software versions
- V07.xx.x
- V08.00.x
- V08.01.x
- V08.02.x
- V08.03.x
- V08.04.x
- V08.05.x
Available parameters and data display
- tcPCO₂
- tcPO₂
*PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected
Sentec Interface Settings
Serial Interface Protocol:
TCB-Protocol
Additional accessories required
MIB Connector, MIB II Converter, Infinity Adapter, Patch cable, Standard RS232 cable (Dräger)
Dräger guides and manuals
Carescape B450, Carescape B650, Carescape B850, Carescape Canvas, Solar 8000 M, Solar 8000 i, Dash 5000:
Required software versions
- Carescape B450: V1 or higher
- Carescape B650: V1 or higher
- Carescape B850: V1 or higher
- Carescape Canvas: V3.3 or higher
- Solar 8000 M: V4A
- Solar 8000 i: V4E
- Dash 5000: V3A
Supported Sentec software versions
- V08.00.x
- V08.01.x
- V08.02.x
- V08.03.x
- V08.04.x
- V08.05.x
Available parameters and data display
- tcPCO₂
- tcPO₂*
- SpO₂
- PR
- Trend data display
*PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected
Sentec Interface Settings
Serial Interface Protocol:
SentecLink
Baud rate:
19200
Additional accessories required
Unity ID, DIDCA SENTEC, standard LAN cables (GE)
GE Healthcare guides and manuals
BeneView T5/T6/T8/T9
BeneVision N12/N15/N17/N19/N22
Required software versions
BeneView T5/T6/T8/T9:
5.25 or higher
BeneVision N12, N15, N17, N19, N22:
V2.0 or higher
Supported Sentec software versions
- V08.00.x
- V08.01.x
- V08.02.x
- V08.03.x
- V08.04.x
- V08.05.x
Available parameters and data display
- tcPCO₂
- tcPO₂*
- SpO₂
- PR
- Trend data display
*PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected
Sentec Interface Settings
Serial Interface Protocol:
SentecLink
Baud rate:
115200
Additional accessories required
BeneLink ID Adapter and Cable C (REF PN-115-011468-00 Mindray)
Mindray guides and manuals
Intellivue
Required software versions
IntelliVue MP 40/50/60/70/80/90 or MX600/700/800 via VueLink Interface:
All SW Versions
IntelliVue MP40/50/60/70/80/90 or MX600/700/750/800/850 with IntelliBridge EC10:
H.15 or higher
IntelliVue MX400/450/500/550 with IntelliBridge EC10:
All SW Versions
Supported Sentec software versions
IntelliVue MP 40/50/60/70/80/90 or MX600/700/800/850 via VueLink Interface:
- V07.xx.x
- V08.00.x
- V08.01.x
- V08.02.x
- V08.03.x
- V08.04.x
- V08.05.x
IntelliVue MP40/50/60/70/80/90 or MX600/700/750/800 with IntelliBridge EC10:
- V07.xx.x
- V08.00.x
- V08.01.x
- V08.02.x
- V08.03.x
- V08.04.x
- V08.05.x
IntelliVue MX400/450/500/550 with IntelliBridge EC10:
- V07.xx.x
- V08.00.x
- V08.01.x
- V08.02.x
- V08.03.x
- V08.04.x
- V08.05.x
Available parameters and data display
- tcPCO₂
- tcPO₂*
- SpO₂
- PR
* PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected.
Sentec Interface Settings
Serial Interface Protocol:
Philips Vuelink/Intellibridge 1/2
Additional accessories required
IntelliBridge EC5 and EC10, LAN cable (Philips)
Philips guides and manuals
Flexport System UltraView SL (90442A-03)
Required software version
V2.00.02EN
Supported Sentec software version
- V07.xx.x
- V08.00.x
- V08.01.x
- V08.02.x
- V08.03.x
- V08.04.x
- V08.05.x
Available parameters and data display
- tcPCO₂
- SpO₂
- PR
- Trend data display
Sentec Interface Settings
Serial Interface Protocol:
Spacelabs Flexport
Additional accessories required
Flexport interface, SDLC cable, SDLC terminator, Transition connector, Modular Cable (Spacelabs)
Spacelabs guides and manuals
Patient Data Management Systems (PDMS)
Medical Device information platform (MDIP)
Required software versions
- Sentec A5.x and higher
Supported Sentec software versions
- V07.xx.x
- V08.00.x
- V08.01.x
- V08.02.x
- V08.03.x
- V08.04.x
- V08.05.x
Available parameters and data display
- tcPCO₂
- tcPO₂*
- SpO₂
- PR
* PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected.
Sentec Interface Settings
Serial Interface Protocol:
SentecLink
Baud rate
115200 or lower
Additional accessories required
Device ID Module (DIM )(Capsule), Standard RS232 and LAN cable
Capsule guides and manuals
INTEGRATED CARE MANAGER (ICM)
Required software versions
- Version 9.2
Supported Sentec software versions
- V08.00.x
- V08.01.x
- V08.02.x
- V08.03.x
- V08.04.x
- V08.05.x
Available parameters and data display
- tcPCO₂
- TcPO₂*
- SpO₂
- PR
* PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected.
Sentec Interface Settings
Serial Interface Protocol:
SentecLink
Baud rate
115200 or lower
Additional accessories required
Standard RS232 cable or USB2Serial converter cable
Dräger guides and manuals
Mirrel
Required software versions
- 1.0.2.0 or higher (driver)
Supported Sentec software versions
- V08.05.x
Available parameters and data display
- tcPCO₂
- TcPO₂*
- SPO₂
- PR
- HP
- PI
- Sensor temperature
*PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected.
Sentec Interface Settings
Serial integration
SentecLink
Baud rate
115200
Regional Availability
Japan Only
Metavision
Required software versions
- V5 or higher
Supported Sentec software versions
- V08.00.x
- V08.01.x
- V08.02.x
- V08.03.x
- V08.04.x
- V08.05.x
Available parameters and data display
- tcPCO₂
- TcPO₂*
- SPO₂
- PR
*PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected.
Sentec Interface Settings
Serial Interface Protocol
SentecLink
Baud rate
115200
Additional accessories required
Digit Connect (iMDSoft), Serial to LAN converter cable 250372-15D, Standard LAN cable
iMDsoft guides and manuals
Prime Gaia
Required software versions
- 03-01 or higher
Supported Sentec software versions
- V08.02.x
- V08.03.x
- V08.04.x
- V08.05.x
Available parameters and data display
- tcPCO₂
- TcPO₂*
- SpO₂
- PR
- HP
- PI
- Sensor temperature
- Alarms and status codes
*PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected.
Sentec Interface Settings
- Serial integration: SentecLink, baud rate 115200 or lower or alternative
- Lan integration
Regional Availability
Japan Only
VISI-DOWNLOAD
Required software versions
- 140715 or higher
Supported Sentec software versions
- V08.00.x
- V08.01.x
- V08.02.x
- V08.03.x
- V08.04.x
- V08.05.x
Available parameters and data display
- tcPCO₂
- tcPCO₂ (drift corrected)
- TcPO₂*
- SpO₂
- PR
- Trend data display
*PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected.
Sentec Interface Settings
Serial Interface Protocol:
SentecLink
Baud rate:
115200 or lower
Additional accessories required
Standard RS232 cable or USB2 Serial converter cable
Stowood guides and manuals
TZM (via) UMG to GE Healthcare’s patient monitors
Required software versions
- V1.0
Supported Sentec software versions
- V08.00.x
- V08.01.x
- V08.02.x
Available parameters and data display
- tcPCO₂
- TcPO₂*
*PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected.
Sentec Interface Settings
Serial Interface Protocol:
SentecLink
Baud rate:
115200
Additional accessories required
UMG
NivolisAnton
Required software versions
- SW-Version A009-074B0-4.4.1 or higher
Supported Sentec software versions
- V08.00.x
- V08.01.x
- V08.02.x
- V08.03.x
- V08.04.x
- V08.05.x
Available parameters and data display
- tcPCO₂
- tcPCO₂ Drift Corrected
- SpO₂
- PR
- PI
- Status codes
- Signal quality
- Download of SDL files for import into V-STATS (trend data display, analysis and reporting)
Sentec Interface Settings
Serial Interface Protocol:
SentecLink
Baud rate:
115200
Additional accessories required
NivolisAnton, screwdriver
Ventilators
VENDOM 30/40
Required software versions
- V 2.05 or higher
Supported Sentec software versions
- V07.xx.x
- V08.00.x
- V08.01.x
- V08.02.x
- V08.03.x
- V08.04.x
- V08.05.x
Available parameters and data display
- tcPCO₂
- SpO₂
- PR
- Trend data display
Sentec Interface Settings
Serial Interface Protocol:
SentecLink
Baud rate:
115200
Additional accessories required
USB2 Serial converter cable
Air Liquide guides and manuals
Vivo 45, Nippy
Required software versions
Vivo 45:
1.0.33 or higher
Vivo 45 LS:
3.0.44 or higher
Vivo 45 L (US):
5.0.01 or higher
Nippy 4+ (UK only):
4.0.51 or higher
Supported Sentec software versions
- V07.xx.x
- V08.00.x
- V08.01.x
- V08.02.x
- V08.03.x
- V08.04.x
Available parameters and data display
- tcPCO₂
- PR
- Trend data display
Sentec Interface Settings
Serial Interface Protocol:
SentecLink
Baud rate:
57600 or lower
Additional accessories required
Serial adapter cable (REF 006179) (Breas)
Breas guides and manuals
EO-150
Required software versions
- V 5.0.0 or higher
Supported Sentec software versions
- V08.00.x
- V08.01.x
- V08.02.x
- V08.03.x
- V08.04.x
- V08.05.x
Available parameters and data display
- tcPCO₂
- tcPCO₂ (drift corrected)
- SpO₂
- PR
- Trend data display
Sentec Interface Settings
Serial Interface Protocol:
SentecLink
Baud rate:
115200
Additional accessories required
Standard serial RS232 to USB converter cable, optional: EOVE Clinical Software
EOVE guides and manuals Download the EOVE guide:
Trilogy, A30, A40
Required software versions
Trilogy Evo, Trilogy Evo O2, Trilogy EV300:
- 1.05.06.00 or higher (US)
- 1.06.06.00 or higher (outside US)
A30 EFL, A40 Pro, A40 EFL:
- 1.1.0.0 and higher:
Supported Sentec software versions
- V08.00.x
- V08.01.x
- V08.02.x
- V08.03.x
- V08.04.x
- V08.05.x
Available parameters and data display
Trilogy Evo, Trilogy Evo O2, Trilogy EV300:
- tcPCO₂
- Trend data display
A30 EFL, A40 Pro, A40 EFL:
- tcPCO₂
- SpO₂
- PR
- Trend data display
Sentec Interface Settings
Trilogy Evo, Trilogy Evo O2, Trilogy EV300:
- Serial Interface Protocol: SentecLink
- Baud rate: 19200
A30 EFL, A40 Pro, A40 EFL:
- Serial Interface Protocol: SentecLink
- Baud rate: 115200
Additional accessories required
Trilogy Evo, Trilogy Evo O2, Trilogy EV300:
Isolated USB-A to DB9 cable (REF 1127405), serial null modem cable (Philips Respironics)
A30 EFL, A40 Pro and A40 EFL:
Transcutaneous CO₂ Connection Cable (REF 1146885), Link Module (REF 1061644A), Philips Null Modem, SD Card (Philips Respironics)
Philips guides and manuals
Monsson 4 via remote display:
Required software versions
- Monsoon 4: 1.0 or higher
- Remote Display: 1.0 or higher
Supported Sentec software versions
- V08.00.x
- V08.01.x
- V08.02.x
- V08.03.x
- V08.04.x
- V08.05.x
Available parameters and data display
- tcPCO₂
- SpO₂
- PR
- Trend data display
Sentec Interface Settings
Serial Interface Protocol:
SentecLink
Baud rate
115200 or lower
Additional accessories required
ACUDis remote display, RJ45 to RS232 cable (Thora Tech)
Thora Tech guides and manuals
Digital Connectivity to PG/PSG
CID102L, CID102LB, CID-LXe, CID-LXe-206d, CID-LX, CID-LX-206d
Required software versions
2.2.4 or higher
Supported software versions
- V07.xx.x
- V08.00.x
- V08.01.x
- V08.02.x
- V08.03.x
- V08.04.x
- V08.05.x
Available parameters and data display
- tcPCO₂
- tcPCO₂ (drift corrected)
- SpO₂
- PR
- Trend data display
Sentec Interface Settings
Serial Interface Protocol
SentecLink
Baud rate
115200
Additional accessories required?
Cidelec Analysis Software, Standard RS232 cable or USB2Serial converter cable
Cidelec guides and manuals
NOX A1, NOX A1s, NOX C1, NOX T3
Required software versions
NOX A1 (via W7Link):
FW 1.1.6 and Noxturnal 5.1.3
NOX A1s (via C1):
FW 3.0.9 and Noxturnal 6.3.1 or higher
NOX C1:
FW 1.1.0 and Noxturnal 6.0.0 or higher
NOX T3 (via W7Link):
FW 1.6.1 and Noxturnal 5.1.3
Supported Sentec software versions
- V08.00.x
- V08.01.x
- V08.02.x
- V08.03.x
- V08.04.x
- V08.05.x
Available parameters and data display
- tcPCO₂
- tcPO₂*
- SpO₂
- PR
- Trend data display
*PO₂ monitoring is only supported if PO₂ channel is enabled and OxiVentTM sensor is connected.
Sentec Interface Settings
Serial Interface Protocol
SentecLink
Baud rate
115200
NOX C1:
LAN integration possible (alternative)
Additional accessories required
NOX A1 or T3 (via W7Link)
W7Link, Mini USB Cable , Noxturnal (Nox)
NOX C1 (serial)
RS-232 to 3.5mm jack serial cable, Standard Ethernet cable, Noxturnal (Nox)
NOX C1 (LAN)
Switch, three standard Ethernet cables, Noxturnal (Nox)
Nox guides and manuals
Analog Connections for the tCOM+ and the Sentec Digital Monitor
- E-Series: PSG Cable E
- Grael: PSG Cable E
- Grael 4k: PSG Cable E
- Siesta: PSG Cable G
- Somte: PSG Cable E
- SleepDoc Porti 7: PSG Cable D
- SleepDoc Porti 8: PSG Cable D
- MiniScreen PRO: PSG Cable D
- MiniScreen plus: PSG Cable D
- Sonata: PSG Cable B
- Comet / Aurora: PSG Cable E
- Comlab: PSG Cable H
- Deltamed Brain Box: PSG Cable J
- Deltamed Duo 44: PSG Cable H
- Embletta Gold: PSG Cable L
- Embletta MPR PG: PSG Cable L
- Embla NDx: PSG Cable F
- Embla SDx: PSG Cable E
- Embla S7000 / N7000: PSG Cable A
- Embla Titanium (Isolation Box): PSG Cable E
- Embla Titanium: PSG Cable K
- Embla Installation Manual
- Mdrive: PSG Cable N
- TX Proxy: PSG Cable N
- PAIB2 interface box: PSG Cable F
- Sleepvirtual: PSG Cable E
- BWII (SI1089)**: PSG Cable E
- BWIII PSG: Cable E
** Product number of accessory cables or optocoupler required for installation in brackets behind the product name
- SOMNOscreen (SSC082)**: PSG Cable Q
- SOMNOscreen plus (SSC082)**: PSG Cable Q
** Product number of accessory cables or optocoupler required for installation in brackets behind the product name
- Visi-3: PSG Cable E
- Visi-System: PSG Cable E
- Alice 3 / 4 / 5 / 6 (square plug): PSG Cable B
- Alice 3 / 4 / 5 / 6 (round plug): PSG Cable F
- SOMNOlab: PSG Cable C
- SOMNOlab2: PSG Cable C
Please refer to the tCOM+ menu ‘Advanced Settings/Interfaces/Analog/PSG/Channel Assignment’ respectively the Sentec Digital Monitor (SDM) menu ‘Interfaces/Analog Outputs’ for ‘Channel-to-Parameter’ assignment.
For instructions to interface the tCOM+ or SDM with polygraphic or polysomnographic systems please refer to the respective manufacturer’s instructions for use, to the tCOM+ IFU (HBQ-179), to the chapter ‘Analog Outputs’ in the Technical Manual for the SDM (HB-005752 – HBQ-11) or the Analog Interface Calibration Guide for the SDM (RF-013071- MMQ-96).
The IFUs and Technical Manual are available as a PDF on the Sentec website at https://www.sentec.com/ifu